Remote monitoring made practical where cost or complexity may otherwise be a barrier.

 

A perfect complement to the sophisticated Acuity Central, providing remote vital sign monitoring wherever needed.

Vital signs displayable

  • Heart Rate + 2 leads of ECG
  • Non-Invasive Blood Pressure
  • Invasive Blood Pressure
  • Pulse Oximetry
  • Impedance Respiration
  • Temperature
  • Capnography

 

Applications

  • Emergency Rooms
  • Cardiac Cath Labs
  • Gastro-Intestinal Endoscopy Departments
  • Intensive and Critical Care Units
  • Post-Anesthesia Care Units

 

Features

  • HIPAA compliant
  • Performs in all patient care areas
  • Provides visual and audible alarms
  • Displays up to seven waveforms per patient
  • Monitor patients in 1, 4, 6 or 8 patient format
  • Print on alarm, tabular and graphical trends
  • PDF output option for EMR system entry

Single Bed Display

Single Bed Display

Eight Bed Display

Eight Bed Display

Specifications

Monitor Compatibility

Model/Software Rev.

Propaq 100 / 8.40.00

Propaq Encore / 1.31.01 to 2.40.00

Propaq CS / 3.00.00 to 3.70.00

Atlas / BB.02.0000

Formats

1, 4, 6 or 8 beds

Time Base

25mm/sec & 50mm/sec, except for CO2 and RESP, which are at 6.25 mm/sec.

Patient/Bed Information

Display Patient Name:

Bed Name:

Print on Alarm:

Pacer:

Yes / No (20 characters max)

5 Characters max

Yes/No

Choice to display pacer flag (Propaq only)

Vital Signs

Waveforms:

Numeric Values:

ECG1, ECG2, RESP, INVP1, INVP2, SPO2, CO2

HR or PR, BP Sys, Dias, and Mean, NIBP, SpO2, EtCO2/InCO2, Resp/Breath Rate, Temperature

Alarms

Visual:

Audible Suspend:

Flashing yellow bell. Alarming numerics flash red.

3 min. when suspended at central station

Control

Keyboard

Video Output

Analog RGB:

Sync:

0.7 Vpp, 75 ohms

TTL Negative going

Resolutions:

1280x1024 (4:3)

1680x1050 (16:10)

1920x1080 (16:9)

Horizontal Rate

64.8 kHz

65.5 kHz

66.7 kHz

Vertical Rate

60.8 Hz

60.2 Hz

59.7 Hz

Printing

Graphic Trend:

Tabular Trend:

Alarm or Manual:

Plot of 8 hr trend history sampled every 2 min.

8 hr of numeric values sampled every 2 min.

A one-hour trend plot precedes the record which is plotted oldest to newest, 12 seconds before and 4 seconds following an alarm or manual initiation.

Physical

Size:

MDP2020-0401

H 2.3” (58 mm)

W 10.2” (259 mm)

D 8.9” (226 mm)

Power Adapter

5V @2A. Certified to be in compliance with the Electrical Safety and

Environmental standards listed below. 100-240V, .38A@100V, .24A@240V.

Weight:
3.0 lbs. (1.36 kg)

AC Input

100-240V, 50-60Hz, .4A at 100V, .25A at 240V

Electrical Safety

Conforms to IEC 60601-1 as Class 1 device for UL 60601-1

Issued: 2003-04-25 Ed.:1 Rev:2006/04/26

CB Scheme testing

Environmental

Temperature:

Operating

0 to 40° C

Storage

-40 to +85° C

Humidity:

5% to 95%, non-condensing

EMC

IEC 60601-1-2:2007 Class B. CISPR 11:2009 (Amended by A1:2010)

 

FDA

Device 510K listed

 

 

Vital signs available for display from each Welch Allyn patient monitor

  Quicksigns Propaq 100 Series Propaq CS & Encore Series Atlas
Waveforms
ECG1 X X X
ECG2 X
P1 X X
P2 X X
SPO2 X X X
CO2 X X X
RESP X X
Numeric Values
Heart Rate X X X X
P1 X X
P2 X X
NIBP X X X X
SPO2 X X X X
CO2 X X X

RR (Respiration Rate)

BR (Breath Rate from CO2)

BR only X X
TEMP T1 T1 T1, T2 T1

Propaq©, Quicksigns©, and Atlas© are registered trademarks of Welch Allyn, Inc.

 

 

Quality System

VIDCO has established and maintains a quality system that conforms with FDA 21 CFR 820 regulation and ISO 13485 standard